THe quantitative determination of sildenafil in human plasma by high performance liquid chromatography - mass spectrometry (lc-ms/ms)
DOI:
https://doi.org/10.15587/2519-4852.2016.87257Keywords:
sildenafil, sildenafil-d8, oral spray, pharmacokinetics, blood plasma, quantification, HPLC-MS/MS, internal standard selection, matrix effect, analytical method validationAbstract
Purpose. For the study of pharmacokinetics of sildenafil oral spray “Tegrum”, developed by pharmaceutical company “Microkhim” (Rubezhnoe, Ukraine) with the aim to improve the pharmacokinetic characteristics of existing solid formulations for oral use it was worked out and validated a rapid, simple and specific method of quantitative determination of sildenafil concentration in human plasma using internal standard labeled with 8 deuterium atoms - sildenafil-d8.
Methods. Sildenafil was extracted from plasma by the liquid-liquid extraction, and quantitatively determined in supernatant by means of а high performance liquid chromatography - tandem mass spectrometry method (HPLC-MS/MS). Separation of components was performed using the chromatographic column Kinetex 2.6u C18, 50 x 4.6 mm, 2.6 μm (Phenomenex) at 45˚C at isocratic elution mode and mobile phase, that consisted of 2 mM aqueous ammonium acetate solution (pH 5,45 ± 0,05) and acetonitrile (10/90, vol./vol.); flow rate equaled to 0.4 ml/min. Ionization of sildenafil and sildenafil-d8 was performed in the positive electrospray mode (ESI, Positive). Multi reactions monitoring (MRM) mode was used for selected daughter and parent ions detection with the following settings of the mass spectrometer: m/z 475,20/283,20 and 483.20/283.20 for sildenafil and sildenafil-d8 respectively.
Results. The method described above was validated in the linear range 10,00-400,00 ng/mL of sildenafil with a correlation coefficient (r2) 0,99764 and demonstrated reliable accuracy and reproducibility of results.
Conclusions. Utilization of elaborated method of fast direct liquid - liquid sildenafil extraction from the blood plasma with simultaneous proteins precipitation, avoiding commonly used supernatant evaporation and subsequent restoration of dry residue at preanalytical stage allowed to achieve target sensitivity of determination and reduced the analysis runtime. Usage of the deuterated internal standard (sildenafil-d8) at analytical stage gave an opportunity to reduce the limit of quantification, to achieve high sensitivity and specificity, as well as suppress matrix effect. Rather high analytical characteristics of the method proposed allow its application in routine bioanalysis during pharmacokinetic studies.
References
- Gorpinchenko, I. I., Miroshnikov, Y. O. (2003). Erectile disfunction. Lviv: Medical world, 88.
- Kanjanawart, S., Kongyingyoes, B., Gaysornsiri, D., Tangsucharit, P., Puapairoj, P., Vannaprasaht, S., Tiamkao, S. et. al. (2008). Bioequivalence and Pharmacokinetic Study of Sildenafil in Healthy Thai Male Volunteers. Srinagarind Medical Journal, 23 (1), 38–44.
- Mahmoudian, M., Falahatpishe, H., Taiebi, L., Moghadam, E., Gholamine, B. (2010). Determination of pharmacokinetic parameters of sildenafil in Iranian volunteers by an HPLC method. Turkish Journal of Pharmaceutical Sciences, 7 (1), 69–74.
- Mahmoudian, M. (2005). Sildenafil Determination in Various Matrices: A Review. Iranian Journal of Pharmacology & Therapeutics, 4 (2), 72–75.
- Dilek, I., Cooper, J., Aijaz, S. (2015). Selection of Internal Standards for LC-MS/MS Applications. Mass Spectrometry Applications to the Clinical Lab (MSACL), 3, 31–42.
- Krishnamurthy, B., Shivaprakasha, R., Amitabha, G. (2009). Quantitation of Sildenafil in Human Plasma by Reverse Phase High Performance Liquid Chromatography. Pharmacology online, 2, 64–74.
- Tripathi, A. S., Sheikh, I., Dewani, A. P., Shelke, P. G., Bakal, R. L., Chandewar, A. V., Mazumder, P. M. (2013). Development and validation of RP-HPLC method for sildenafil citrate in rat plasma-application to pharmacokinetic studies. Saudi Pharmaceutical Journal, 21 (3), 317–321. doi: 10.1016/j.jsps.2012.09.003
- Al-Ghazawi, M., Tutunji, M., AbuRuz, S. (2007). Simultaneous determination of sildenafil and N-desmethyl sildenafil in human plasma by high-performance liquid chromatography method using electrochemical detection with application to a pharmacokinetic study. Journal of Pharmaceutical and Biomedical Analysis, 43 (2), 613–618. doi: 10.1016/j.jpba.2006.07.028
- Berzas, J. J., Rodriguez, J., Villasenor, M. J., Contento, A. M., Cabello, M. P. (2002). Validation of a capillary gas chromatographic method for the determination of Sildenafil Citrate in its pharmaceutical formulations (Viagra). Experimental design for evaluating the ruggedness of the method. Chromatographia, 55 (9-10), 601–606. doi: 10.1007/bf02492908
- Marcelin-Jimenez, G., Angeles-Moreno, A. P., Contreras-Zavala, L., Garcia-Gonzalez, A., Ramirez-San Juan, E. (2012). Comparison of Fasting Bioavailability Among 100-mg Commercial, 100-mg Generic, and 50-mg Chewable Generic Sildenafil Tablets in Healthy Male Mexican Volunteers: A Single-Dose, 3-Period, Crossover Study. Clinical Therapeutics, 34 (3), 689–698. doi: 10.1016/j.clinthera.2012.01.021
- Alkharfy, K. M. (2009). Simple and sensitive LC-ESI-MS method for the quantitation of sildenafil in plasma samples. Journal of Separation Science, 32 (22), 3866–3870. doi: 10.1002/jssc.200900469
- Zayed, R., Kamel, A. O., Shukr, M., El-Shamy, A. E.-H. (2012). An in vitro and in vivo comparative study of directly compressed solid dispersions and freeze dried sildenafil citrate sublingual tablets for management of pulmonary arterial hypertension. Acta Pharmaceutica, 62 (3), 411–432. doi: 10.2478/v10007-012-0027-9
- Challa, B. R., Awen, B. Z., Chandu, B. R., Khagga, M., Bannoth, C. K., Kanala, K. et. al. (2010). Sildenafil and N-desmethyl sildenafil quantification in human plasma by HPLC coupled with ESI-MS/MS detection: Application to bioequivalence study. Analytical Methods, 2 (8), 1043–1050. doi: 10.1039/c0ay00062k
- Johnson, R. D., Lewis, R. J. (2006). Identification of Sildenafil (Viagra®) and Its Metabolite (UK-103,320) in Six Aviation Fatalities. Federal Aviation Administration. Washington, 11.
- Sychev, K. S. (2010). Practical guidance on liquid chromatography. Moscow: Tehnosfera, 272.
- Good Laboratory Practice. Oecd principles and guidance for compliance monitoring (2005). OECD.
- On approval of documents on quality assurance of medicines (2009). The Ministry of Health of Ukraine, # 95.
- Bioanalytical method validation: Guidance for industry (2001). U. S. Department of Health and Human Services, Food and Drug Administration. Center for Drug Evalution and Research (CDER). Center for Veterinary Medicine (CVM).
- Guideline on bioanalytical method validation (2011). European Medicines Agency (EMEA/CHMP/EWP/192217/2009).
- Zhukova, N. A., Libina, V. V., Kudris, I. V., Padalko, N. N. (2013). Bioanalytical Method Validation. Kyiv, 35.