Methods of improving the reconstructed image in systems with coded apertures

Authors

  • Александр Анатольевич Никуляк Scientific Research and Desing Institute "Iskra" Zveyneka st., 145с, Lugansk, 91033, Ukraine

DOI:

https://doi.org/10.15587/1729-4061.2012.5249

Keywords:

coded apertures, decoding, software simulation modeling

Abstract

The application of systems with a coded aperture to obtain tomographic information has its pros and cons. The advantages of such systems are the high sensitivity and noise-immunity (compared, for example, with tomographs on the basis of gamma-cameras), the ability to focus on a given surface within the object, etc. The drawbacks include the near-field artifacts, which occur during the decoding of the source position and depend on the choice of a coding mask and geometric parameters of the system. In addition, the quality of the restored image is affected by the accuracy of the formation of the matrix of countability. The article studies the impact of the interaction of gamma-radiation with a substance on the quality of the formation of the matrix of countability and, as a result, on the quality of the restored image. It describes the method of formation of the correcting processing array and its use as a way to improve the restored image

Author Biography

Александр Анатольевич Никуляк, Scientific Research and Desing Institute "Iskra" Zveyneka st., 145с, Lugansk, 91033

Junior research fellow

References

  1. Fenimore E.E. “Coded aperture imaging with uniformly redundant arrays” [Текст] / E.E. Fenimore, T.M. Cannon // Applied Optics – 1978. – T. 17, № 3. – C. 337 – 347.
  2. Fenimore E.E. “Uniformly redundant arrays: digital reconstruction methods”[Текст] / E.E. Fenimore, T.M. Cannon // Applied Optics – 1981. – T. 20, № 10. – C. 1858 – 1864.
  3. Liu Y.H. “A novel geometry for SPECT imaging associated with the EM-type blind deconvolution method” [Текст] / Y.H. Liu, A. Rangarajan, D. Gagnon, M. Therrien, A.J. Sinusas, F.J.Th. Wackers, I.G. Zubal // IEEE Transactions on Nuclear Science – 1998. – T. NS45 № 4. – C. 2095 – 2101.
  4. Koral K.F. “Application of ART to Time-coded Emission-Tomography” [Текст] / K.F. Koral, W.L. Rogers // Physics in Medicine and Biology – 1979. T. 24, №5. – C. 879 – 894.
  5. Fleming J.S. “An evaluation of techniques for stationary coded aperture three-dimensional imaging in nuclear medicine” [Текст] / J.S. Fleming, B.A. Goddard // Nuclear Instruments and Methods in Physics Research – 1984. T. 221. C. – 242 – 246.
  6. Geant4: a toolkit for the simulation of the passage of particles through matter [Internet resource] / Available at: http://geant4.cern.ch.

Published

2012-12-04

How to Cite

Никуляк, А. А. (2012). Methods of improving the reconstructed image in systems with coded apertures. Eastern-European Journal of Enterprise Technologies, 6(9(60), 37–40. https://doi.org/10.15587/1729-4061.2012.5249

Issue

Section

Information and controlling system