Designing a bench for testing medically and technically an information and measuring system for multi-frequency electrical impedance tomography of human lungs

Authors

DOI:

https://doi.org/10.15587/1729-4061.2021.237529

Keywords:

electrical impedance tomography, lungs, perfusion, information and measuring system, bench, visualization, conductivity

Abstract

This paper solves the problem of developing and creating multifunctional tools for conducting research into systems of multi-frequency electrical impedance tomography of human lungs. A test bench consisting of modern medical equipment, simulating pre-and postoperative environments to create conditions that are most similar to clinical ones, under which it is planned to operate multi-frequency electrical impedance tomography systems, was proposed and manufactured. This makes it possible to reduce significantly the time for approbation, testing, and elimination of practical inaccuracies and problems of clinical application of the developed medical and technical facilities. A positive result is achieved due to the possibility of forming new test plans with specified conditions and different levels of complexity. This enables enhancing the effectiveness of subsequent clinical tests on patients who are treated in a resuscitation unit or an intensive care unit. The operability of the bench is proved by the repeatability of obtained results of monitoring ventilation and perfusion for each examined person, the continuity of dynamic visualization of the breathing process, as well as a high degree of correlation of obtained values of differences of potentials with the readings of a bedside monitor of a patient. An information and measuring system of multi-frequency electric impedance tomography of human lungs, developed by the author earlier, was used as the EIT device. The EIT tests were performed for the frequency range of 50 kHz – 400 kHz at a current of 5 mA. All experimental studies involved volunteers who gave written information consent to participate in the tests. The results of the research show that the proposed bench can be used in practice to solve a wide range of scientific and applied problems in the field of electrical impedance tomography

Author Biography

Grayr Aleksanyan, Platov South-Russian State Polytechnic University (NPI)

PhD

Department «Information and Measurement Systems and Technologies»

References

  1. Pekker, Ya. S., Brazovskiy, K. S., Usov, V. N. (2004). Electrical impedance tomography. Tomsk: NTL, 192.
  2. Abboud, M., Guardo, R., Martineau, R., Taillefer, Pelletier, J. C. (1995). Monitoring of peripheral edema using electrical bioimpedance measurements. EMBS, Conf IEEE, 641–642. doi: http://doi.org/10.1109/iembs.1995.575290
  3. Pomprapa, A., Schwaiberger, D., Pickerodt, P., Tjarks, O., Lachmann, B., & Leonhardt, S. (2014). Automatic protective ventilation using the ARDSNet protocol with the additional monitoring of electrical impedance tomography. Critical Care, 18 (3), R128. doi: http://doi.org/10.1186/cc13937
  4. Gorbatenko, N. I., Katsupeev, A. A., Aleksanyan, G. K., Shcherbakov, I. D., Kucher, A. I. (2020). Integration principles of the electrical impedance tomography module with a lung ventilator. Journal of Physics: Conference Series, 1679, 032080. doi: http://doi.org/10.1088/1742-6596/1679/3/032080
  5. Leonhardt, S., Lachmann, B. (2012). Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring? Intensive Care Medicine, 38 (12), 1917–1929. doi: http://doi.org/10.1007/s00134-012-2684-z
  6. Zhao, Z., Fu, F., Frerichs, I. (2020). Thoracic electrical impedance tomography in Chinese hospitals: a review of clinical research and daily applications. Physiological Measurement, 41 (4), 04TR01. doi: http://doi.org/10.1088/1361-6579/ab81df
  7. Becher, T., Buchholz, V., Hassel, D., Meinel, T., Schädler, D., Frerichs, I., Weiler, N. (2021). Individualization of PEEP and tidal volume in ARDS patients with electrical impedance tomography: a pilot feasibility study. Annals of Intensive Care, 11 (1). doi: http://doi.org/10.1186/s13613-021-00877-7
  8. Spinelli, E., Kircher, M., Stender, B., Ottaviani, I., Basile, M. C., Marongiu, I. et. al. (2021). Unmatched ventilation and perfusion measured by electrical impedance tomography predicts the outcome of ARDS. Critical Care, 25 (1). doi: http://doi.org/10.1186/s13054-021-03615-4
  9. Wolf, G. K., Gómez-Laberge, C., Rettig, J. S., Vargas, S. O., Smallwood, C. D., Prabhu, S. P. et. al. (2013). Mechanical Ventilation Guided by Electrical Impedance Tomography in Experimental Acute Lung Injury*. Critical Care Medicine, 41 (5), 1296–1304. doi: http://doi.org/10.1097/ccm.0b013e3182771516
  10. Wolf, G. K., Gómez-Laberge, C., Rettig, J. S., Vargas, S. O., Smallwood, C. D., Prabhu, S. P. et. al. (2013). Mechanical Ventilation Guided by Electrical Impedance Tomography in Experimental Acute Lung Injury*. Critical Care Medicine, 41 (5), 1296–1304. doi: http://doi.org/10.1097/ccm.0b013e3182771516
  11. Muders, T., Luepschen, H., Zinserling, J., Greschus, S., Fimmers, R., Guenther, U. et. al. (2012). Tidal recruitment assessed by electrical impedance tomography and computed tomography in a porcine model of lung injury*. Critical Care Medicine, 40 (3), 903–911. doi: http://doi.org/10.1097/ccm.0b013e318236f452
  12. Nestler, C., Simon, P., Petroff, D., Hammermüller, S., Kamrath, D., Wolf, S. et. al. (2017). Individualized positive end-expiratory pressure in obese patients during general anaesthesia: a randomized controlled clinical trial using electrical impedance tomography. British Journal of Anaesthesia, 119 (6), 1194–1205. doi: http://doi.org/10.1093/bja/aex192
  13. Mauri, T., Eronia, N., Turrini, C., Battistini, M., Grasselli, G., Rona, R. et. al. (2016). Bedside assessment of the effects of positive end-expiratory pressure on lung inflation and recruitment by the helium dilution technique and electrical impedance tomography. Intensive Care Medicine, 42 (10), 1576–1587. doi: http://doi.org/10.1007/s00134-016-4467-4
  14. Kim, M., Bae, J., Yoo, H.-J. (2017). Wearable 3D lung ventilation monitoring system with multi frequency electrical impedance tomography. 2017 IEEE Biomedical Circuits and Systems Conference (BioCAS). doi: http://doi.org/10.1109/biocas.2017.8325163
  15. Aguiar Santos, S., Czaplik, M., Orschulik, J., Hochhausen, N., Leonhardt, S. (2018). Lung pathologies analyzed with multi-frequency electrical impedance tomography: Pilot animal study. Respiratory Physiology & Neurobiology, 254, 1–9. doi: http://doi.org/10.1016/j.resp.2018.03.016

Downloads

Published

2021-08-31

How to Cite

Aleksanyan, G. (2021). Designing a bench for testing medically and technically an information and measuring system for multi-frequency electrical impedance tomography of human lungs . Eastern-European Journal of Enterprise Technologies, 4(5(112), 27–40. https://doi.org/10.15587/1729-4061.2021.237529

Issue

Section

Applied physics