Exploring dependence of the patints’ radiation dose on the form of power supply voltage of x-ray tube
DOI:
https://doi.org/10.15587/1729-4061.2016.81305Keywords:
numerical modeling, equivalent doses, radiation exposure, medical X-ray examinations, power supply sourcesAbstract
A brief analysis was conducted of organizational and technical methods of reducing the patient’s dose during radiographic examinations that are used in contemporary medicine. The paper considers optimization of the form of power supply voltage of X-ray tube as one of the efficient, but not sufficiently explored, areas of solving this problem. For its analysis, we propose a method of numerical simulation of the process of X-ray screening based on the experimentally obtained data. The results of experimental research are presented.
Authors described the methods and results of calculation of influence of the form of power supply voltage of X-ray emitter on the total radiation dose of patients during medical examinations.
To model the performance of several types of power units that operate on different physical principles, we created mathematical models that made it possible to calculate the instantaneous and integral values of all electrical parameters of X-ray tube, to find the integral dose of radiation in the plane of patient's location and to simulate a generalized reaction of the X-ray receiver of images on the ray flow that runs through the object of examination.
Comparative analysis of the obtained results demonstrated that the total dose of radiation is significantly affected by the design and principle of operation of the power supply unit of X-ray tube, which is a part of the X-ray device. At the equivalent diagnostic quality of images, the difference in the X-ray dose of received by the patient in some modes may exceed 20 %.
The assessment we performed indicates the feasibility of modernization of the outdated X-ray equipment with low frequency power units for the purpose of reducing the radiation dose through the improvement in the form of high voltage.
References
- Order of Ministry of Healh of Ukraine, May 17, 2008, 254 Stat. Instructions on the X-ray examinations frequency of the chest cavity of the certain categories of population in Ukraine. Registered by the Ministry of Justice of Ukraine, June 12, 2008, 524/15215 Stat. Official Bulletin of Ukraine, 1475 Stat, 43372/2008 act code, 45, 41.
- Law of Ukraine, Oct.16, 2012, 5451–VI Stat. The approval of the National Programme on TB for 2012–2016 (2013). Tuberculosis, pulmonary diseases, HIV infection, 2, 85–109.
- Ilina, L., Yarmonenko, S. (Eds.) (2002). Sources and Effects of Ionizing Radiation. Report of the United Nations Scientific Committee on the Effects of Atomic Radiation 2000 the UN General Assembly, with Scientific Annexes, 1 (Sources 1). Moscow, 308.
- Murashko, V., Mechev, D. et al. (2013). Radiation Hygiene. A textbook for medical interns and medical students. Vinnitsa: A New Book, 376.
- Resolution of the Board of State Nuclear Regulation (2010). Situation and problems of radiation protection of staff and patients in the application of ionizing radiation in medicine. Available at: http://www.snrc.gov.ua/nuclear/uk/publish/article/133546
- Andria, G., Attivissimo, F., Di Nisio, A., Lanzolla, A. M. L., Guglielmi, G., Terlizzi, R. (2014). Dose Optimization in Chest Radiography: System and Model Characterization via Experimental Investigation. IEEE Transactions on Instrumentation and Measurement, 63 (5), 1163–1170. doi: 10.1109/tim.2013.2282411
- Berkhout, W. E. (2015). The ALARA-principle. Backgrounds and enforcement in dental practices. Nederlands Tijdschrift voor Tandheelkunde, 122 (05), 263–270. doi: 10.5177/ntvt.2015.5.14227
- Xiao, L., Li, C., Wu, Z., Wang, T. (2016). An enhancement method for X-ray image via fuzzy noise removal and homomorphic filtering. Neurocomputing, 195, 56–64. doi: 10.1016/j.neucom.2015.08.113
- Voisin, P. (2015). Standards in biological dosimetry: A requirement to perform an appropriate dose assessment. Mutation Research/Genetic Toxicology and Environmental Mutagenesis, 793, 115–122. doi: 10.1016/j.mrgentox.2015.06.012
- Andria, G., Attivissimo, F., Guglielmi, G., Lanzolla, A. M. L., Maiorana, A., Mangiantini, M. (2016). Towards patient dose optimization in digital radiography. Measurement, 79, 331–338. doi: 10.1016/j.measurement.2015.08.015
- Stadnyk, L., Shalepa, O., Nosik, O. (2014) Assessment of patient's doses for radiography and their optimization by the establishment of the national diagnostic reference levels. Radiation hygiene, 7 (4), 84–88.
- Seeram, E., Davidson, R., Bushong, S., Swan, H. (2016). Optimizing the Exposure Indicator as a Dose Management Strategy in Computed Radiography. Radiol Technol, 87 (4), 380–391.
- Jia, X., Yan, H., Gu, X., Jiang, S. B. (2012). Fast Monte Carlo simulation for patient-specific CT/CBCT imaging dose calculation. Physics in Medicine and Biology, 57 (3), 577–590. doi: 10.1088/0031-9155/57/3/577
- Lee, C., Lee, C., Staton, R. J., Hintenlang, D. E., Arreola, M. M., Williams, J. L., Bolch, W. E. (2007). Organ and effective doses in pediatric patients undergoing helical multislice computed tomography examination. Medical Physics, 34 (5), 1858–1873. doi: 10.1118/1.2723885
- Alves, A. F. F., Miranda, J. R. de A., Bacchim Neto, F. A., Duarte, S. B., Pina, D. R. de (2015). Construction of pediatric homogeneous phantoms for optimization of chest and skull radiographs. European Journal of Radiology, 84 (8), 1579–1585. doi: 10.1016/j.ejrad.2015.05.015
- Pina, D. R., Duarte, S. B., Morceli, J., Ghilardi Netto, T. (2006). Development of phantom for radiographic image optimization of standard patient in the lateral view of chest and skull examination. Applied Radiation and Isotopes, 64 (12), 1623–1630. doi: 10.1016/j.apradiso.2006.05.006
- Hvay, A., Averyanova, L. (2013). The analysis of the methods for determining the radiation dose in X-ray computed tomography. Bulletin of NTU "KhPI", 39 (1012), 41–47.
- Fujii, K., Nomura, K., Muramatsu, Y., Obara, S., Akahane, K., Kusumoto, M. (2016). Organ Dose Evaluations Based on Monte Carlo Simulation for CT Examinations Using Tube Current Modulation. Radiation Protection Dosimetry. doi: 10.1093/rpd/ncw144
- Rehani, M. M. (2015). Looking into future: challenges in radiation protection in medicine. Radiation Protection Dosimetry, 165 (1-4), 3–6. doi: 10.1093/rpd/ncv071
- Ofori, K., Gordon, S. W., Akrobortu, E., Ampene, A. A., Darko, E. O. (2014). Estimation of adult patient doses for selected X-ray diagnostic examinations. Journal of Radiation Research and Applied Sciences, 7 (4), 459–462. doi: 10.1016/j.jrras.2014.08.003
- Aslamova, L., Golinko-Bezshyyko, L., Solodovnyk, K. (2008) Recommendations for reducing doses to patients by the medical irradiation. Nuclear and Radiation Safety, 4, 77–80.
- Martin, C. J. (2011). Management of patient dose in radiology in the UK. Radiation Protection Dosimetry, 147 (3), 355–372. doi: 10.1093/rpd/ncr386
- Sidnyev, O., Perevoznikov, O. (2009). Exposure of the population in modern medical X-Ray diagnostics and ways to reduce it. Problems of nuclear power and Chernobyl, 11, 146–151.
- Stadnyk, L., Shalopa, O., Nosik, O. (2015). Optimization of doses to patients during X-ray chest. Radiation diagnostics, radiation therapy, 1, 78–81.
- Anyschyk, S., Kovalenko, N., Popov, V., Rahymzhanova, R., Suslin, V. (2008). Optimization of regimes and patient doses for the adequate radiographic information. Clinic: Professional magazine for managers and doctors of all specialties of health facilities in Russia, 6 (1), 30–38.
- Liashchenko, V., Reva, S., Sokol, V., Turchin, O. (2005) A pulse-width control of the X-ray tube anode voltage. Bulletin of Kharkiv National University. Mathematical modeling. Information Technology. Automated control systems, 661, 174–181.
- Dikan, O. (2000). Ray of the light in the X-Ray diagnostics darkness. Mirror of the week, 29 (302).
- Pinchuk, I., Kalinichenko, V. (2011). The inspection experience of X-ray diagnostics devices in medical institutions and personal protective equipment for compliance to the State Standard in Kharkov region as the example. Radiology Magazine, 3, 354–356.
- Laboratory of Intelligent Electronic Systems (2005–2011). The X-ray high frequency power supply IEC-F7: general information. Available at: http://iec-lab.com.ua/group.php?id=8&cid=6
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2016 Sergiy Reva, Maryna Malakhova
This work is licensed under a Creative Commons Attribution 4.0 International License.
The consolidation and conditions for the transfer of copyright (identification of authorship) is carried out in the License Agreement. In particular, the authors reserve the right to the authorship of their manuscript and transfer the first publication of this work to the journal under the terms of the Creative Commons CC BY license. At the same time, they have the right to conclude on their own additional agreements concerning the non-exclusive distribution of the work in the form in which it was published by this journal, but provided that the link to the first publication of the article in this journal is preserved.
A license agreement is a document in which the author warrants that he/she owns all copyright for the work (manuscript, article, etc.).
The authors, signing the License Agreement with TECHNOLOGY CENTER PC, have all rights to the further use of their work, provided that they link to our edition in which the work was published.
According to the terms of the License Agreement, the Publisher TECHNOLOGY CENTER PC does not take away your copyrights and receives permission from the authors to use and dissemination of the publication through the world's scientific resources (own electronic resources, scientometric databases, repositories, libraries, etc.).
In the absence of a signed License Agreement or in the absence of this agreement of identifiers allowing to identify the identity of the author, the editors have no right to work with the manuscript.
It is important to remember that there is another type of agreement between authors and publishers – when copyright is transferred from the authors to the publisher. In this case, the authors lose ownership of their work and may not use it in any way.