Assessing the potential accuracy of a small-sized goniometer with extended dynamic range based on nuclear magnetic resonance
DOI:
https://doi.org/10.15587/1729-4061.2024.313878Keywords:
goniometer, nuclear magnetic resonance, gas cell, optical pumping, Helmholtz coil, Larmor frequencyAbstract
This paper evaluates potential accuracy characteristics of a small-sized goniometer based on nuclear magnetic resonance with an extended dynamic range. This required constructing a model of goniometer errors, estimating its accuracy based on this model, and formulating practical recommendations for the design of such a device based on the accuracy assessment.
To evaluate the accuracy of a nuclear goniometer, a theoretical model was built that makes it possible to determine the optimal operating parameters of the cell gas mixture, the ranges of their permissible changes, the sensitivity of the goniometer, and the dependence of its characteristics on external and internal factors. In particular, the dependence of output signal of the device on the parameters of gas mixture and optical pumping has been determined. For a goniometer with a cell volume of 8 cm3, the optimum temperature is 130 °C, and the optimum intensity of the pumping radiation is 5 mW.
The dependence of output signal on the measured angle of rotation was also established, as well as the noise and error dependence of the device on the permissible values of its parameters. Based on the model built, parameters of a goniometer with a cell volume of 8 cm3 were determined; the maximum angular sensitivity of such a goniometer with complete suppression of technical noise is δφsen=1.0 arcsec. The greatest contribution to the angle measurement error is from the instability of pumping power Ip (ΔIp/Ip=0.05) – 85 %, magnetic field B0 (ΔB0/B0=10-8) – 13 %, temperature T (ΔT/T=0,1) – 2 %.
The goniometer under consideration corresponds to the medium accuracy class, δφtot≥10 arcsec. It could be used in optical manufacturing for operational control, calibration, and certification of optical products. To improve the angular accuracy of the goniometer, it is necessary to increase the stability of the laser pumping intensity
References
- Zheng, L., Tang, Q., Ma, X., Zhang, Y. (1996). High-precision static and dynamic angular measurements with a ring laser gyroю Automated Optical Inspection for Industry, 2899, 50–53. https://doi.org/10.1117/12.253053
- Pisani, M., Astrua, M., Santiano, M., Beverini, N., Di Virgilio, A., Maccioni, E. et al. (2018). G-LAS: a ring laser gyroscope for high accuracy angle measurements. Journal of Physics: Conference Series, 1065, 032009. https://doi.org/10.1088/1742-6596/1065/3/032009
- Wie, Z., Yao, H., Ke, L., Jin, Z., Zi, X. (2022). Calibration and Measurement Method of Laser Gyro Goniometer. Metrology Science and Technology, 66 (4), 40–47.
- Cherepanska, I. Yu., Bezvesilna, O. M., Sazonov, A. Yu. (2019). Precise Intelligent Goniometric System. Visnyk of Vinnytsia Politechnical Institute, 143 (2), 7–14. https://doi.org/10.31649/1997-9266-2019-143-2-7-14
- Zou, W., Huang, Y., Lin, H., Xue, Z. (2024). New Application and Research of Ring Laser Gyroscope in the Field of Angle Metrology. IEEE Transactions on Instrumentation and Measurement, 73, 1–12. https://doi.org/10.1109/tim.2024.3449940
- Mou, J., Pang, B., Huang, T., Ying, G., Shu, X. (2019). A new method to eliminate the misalignment angle in dynamic goniometer based on fiber optic gyro. Optik, 193, 162998. https://doi.org/10.1016/j.ijleo.2019.162998
- Pang, B., Ying, G., Xue, F., Huang, T., Che, S., Shu, X., Mou, J. (2019). Uncertainty analysis of dynamic goniometer based on fiber optic gyroscope. 9th International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test, Measurement Technology, and Equipment, 20, 115. https://doi.org/10.1117/12.2509290
- Ivanov, S. V. (2016). Porivnialnyi analiz efektyvnosti vydiv namotky volokna chutlyvoho elementa volokonno-optychnoho hiroskopa v umovakh zminy temperatury. Naukovi visti natsionalnoho tekhnichnoho universytetu Ukrainy "Kyivskyi politekhnichnyi instytut", 1, 99–106.
- Ivanov, S. V., Muravov, V. V., Oliinyk, P. B. (2024). Pat. No. 156304 UA. Honiometr. MPK G01B11/26; No. u202305588; declareted: 21.11.2023; published: 05.06.2024, Bul. No. 23.
- Skinner, J. G., Ranta, K., Whiting, N., Coffey, A. M., Nikolaou, P., Rosen, M. S. et al. (2020). High Xe density, high photon flux, stopped-flow spin-exchange optical pumping: Simulations versus experiments. Journal of Magnetic Resonance, 312, 106686. https://doi.org/10.1016/j.jmr.2020.106686
- Cates, G. D., Fitzgerald, R. J., Barton, A. S., Bogorad, P., Gatzke, M., Newbury, N. R., Saam, B. (1992). Rb–129Xe spin-exchange rates due to binary and three-body collisions at high Xe pressures. Physical Review A, 45 (7), 4631–4639. https://doi.org/10.1103/physreva.45.4631
- Li, R., Quan, W., Fang, J. (2017). Polarization Measurement of Cs Using the Pump Laser Beam. IEEE Photonics Journal, 9 (6), 1–8. https://doi.org/10.1109/jphot.2017.2761779
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2024 Sergiy Ivanov
This work is licensed under a Creative Commons Attribution 4.0 International License.
The consolidation and conditions for the transfer of copyright (identification of authorship) is carried out in the License Agreement. In particular, the authors reserve the right to the authorship of their manuscript and transfer the first publication of this work to the journal under the terms of the Creative Commons CC BY license. At the same time, they have the right to conclude on their own additional agreements concerning the non-exclusive distribution of the work in the form in which it was published by this journal, but provided that the link to the first publication of the article in this journal is preserved.
A license agreement is a document in which the author warrants that he/she owns all copyright for the work (manuscript, article, etc.).
The authors, signing the License Agreement with TECHNOLOGY CENTER PC, have all rights to the further use of their work, provided that they link to our edition in which the work was published.
According to the terms of the License Agreement, the Publisher TECHNOLOGY CENTER PC does not take away your copyrights and receives permission from the authors to use and dissemination of the publication through the world's scientific resources (own electronic resources, scientometric databases, repositories, libraries, etc.).
In the absence of a signed License Agreement or in the absence of this agreement of identifiers allowing to identify the identity of the author, the editors have no right to work with the manuscript.
It is important to remember that there is another type of agreement between authors and publishers – when copyright is transferred from the authors to the publisher. In this case, the authors lose ownership of their work and may not use it in any way.